Energy calibration of the NaI(Tl) calorimeter of the SND detector using $e^+e^- \to e^+e^-$ events
arXiv:physics/9806014 · doi:10.1016/S0168-9002(98)00348-9
Abstract
Calibration of the three layer NaI(Tl) spherical calorimeter of the SND detector using electron -- positron scattering events is described. Energy resolution of $5 % (FWHM/2.36)$ for 500 MeV photons was achieved.
12 pages, Latex, 8 figures (.EPS)