arXiv:nucl-ex/9510002 · doi:10.1016/0168-9002(95)01015-7

Abstract

A procedure has been developed for the charge, mass and energy calibration of ions produced in nuclear heavy ion reactions. The charge and mass identification are based on a $Δ$E-E technique. A computer code determines the conversion from ADC channels into energy values, atomic number and mass of the detected fragments by comparing with energy loss calculations through a minimization routine. The procedure does not need prior measurements with beams of known energy and charge. An application of this technique to the calibration of the MULTICS apparatus is described.

9 pages, Tex file, 3 postscript figures available upon request from mastinu@bologna.infn.it; to appear in Nucl. Inst. Meth