NewEvery arXiv paper, its researchers & institutions — mapped.
paper

A high precision calibration of the nonlinear energy response at Daya Bay

arXiv:1902.08241

Abstract

A high precision calibration of the nonlinearity in the energy response of the Daya Bay Reactor Neutrino Experiment's antineutrino detectors is presented in detail. The energy nonlinearity originates from the particle-dependent light yield of the scintillator and charge-dependent electronics response. The nonlinearity model is constrained by $γ$ calibration points from deployed and naturally occurring radioactive sources, the $β$ spectrum from $^{12}$B decays, and a direct measurement of the electronics nonlinearity with a new flash analog-to-digital converter readout system. Less than 0.5% uncertainty in the energy nonlinearity calibration is achieved for positrons of kinetic energies greater than 1 MeV.

17 pages, 22 figures, 4 tables. Final version to be published in NIM-A