In-situ Instrumental Setup for Influence Study of Hard-axis Bias Magnetic Field on MR transfer curves of sing MTJ sensor and MTJs array sensor
arXiv:1604.07514
Abstract
Establishment of home-made measurement setups for the characterization of MR sensor is proposed and described here. The MR loops of MR sensors can be obtained with the instrument using two-point probe measurement and four-point probe measurement. Two pairs of Helmholtz coils can supply a hard-axis magnetic field and a soft-axis magnetic field for the sensor. The single MTJ sensor and MTJs array sensor in Wheatstone bridge were characterized and compared here. The influence of hard-bias magnetic field on MR transfer curves of sing MTJ sensor and MTJs array sensor are investigated. The corresponding optimal hard-axis magnetic fields were obtained through Helmholtz coils to eliminate the hysteresis for linear response of single MTJ sensor and MTJs array sensor.