Measurement of the temperature dependence of pulse lengths in an n-type germanium detector
arXiv:1112.5033 · doi:10.1051/epjap/2011110110
Abstract
The temperature dependence of the pulse length was measured for an 18-fold segmented n-type germanium detector in the temperature range of 77-120 K. The interactions of 122 keV photons originating from a Europium-152 source were selected and pulses as observed on the core and segment electrodes were studied. In both cases, the temperature dependence can be well described by a Boltzmann-like ansatz.
17 pages, 2 tables, 13 figures, published in EPJ AP