NewEvery arXiv paper, its researchers & institutions — mapped.
paper

Cryogenic environment and performance for testing the Planck radiometers

arXiv:1001.4644 · doi:10.1088/1748-0221/4/12/T12015

Abstract

This paper is part of the Prelaunch status LFI papers published on JINST: http://www.iop.org/EJ/journal/-page=extra.proc5/jinst The Planck LFI Radiometer Chain Assemblies (RCAs) have been calibrated in two dedicated cryogenic facilities. In this paper the facilities and the related instrumentation are described. The main satellite thermal interfaces for the single chains have to be reproduced and stability requirements have to be satisfied. Setup design, problems occurred and improving solutions implemented are discussed. Performance of the cryogenic setup are reported.

This is an author-created, un-copyedited version of an article accepted for publication in JINST. IOP Publishing Ltd is not responsible for any errors or omissions in this version of the manuscript or any version derived from it. The definitive publisher authenticated version is available online at 10.1088/1748-0221/4/12/T12015