Energy resolution improvement in room-temperature CZT detectors
arXiv:0709.3481 · doi:10.1088/1748-0221/2/12/P12003
Abstract
We present methods to improve the energy resolution of single-channel, room-temperature Cadmium-Zinc-Telluride (CZT) detectors. A new preamplifier design enables the acquisition of the actual transient current from the crystals and straightforward data analysis methods yield unprecedented energy resolution for our test-detectors. These consist of an eV-CAPture Plus crystal as standard and 1 cm cube Frisch collar crystals created in-house from low-grade coplanar grid detectors. Energy resolutions of 1.9% for our collar detectors and 0.8% for the eV crystal at 662 keV were obtained. The latter compares favourably to the best existing energy resolution results from pixel detectors.
12 pages, 10 figures, 1 table, submitted to Journal of Instrumentation